The impact of artificial light scatter on scanning laser tomography
Mark A Burke1, Chaten J Khanna, Angela Miller
1School of Optometry, University of Waterloo, Waterloo, Ontario, Canada.
Purpose:
The impact of cataract (which frequently occurs alongside glaucoma) on scanning laser tomography (SLT) is poorly understood. The aim of this pilot study was to determine the impact of artificial light scatter on SLT estimates of optic nerve head (ONH) topography.
Methods:
The sample comprised 10 healthy, young subjects of mean age 23.5 years. One eye of each subject was randomly selected. Cells filled with increasing concentrations of 0.50-microm diameter polystyrene microspheres were prepared. The cells were mounted in front of the objective lens of the Heidelberg Retina Tomograph (HRT) II and were tilted at an angle of 20 degrees to eradicate any surface reflections. Three sets of ONH scans were initially acquired without any light scatter cell in place and then three further sets were acquired for each of four different concentrations of microspheres in a randomized order. The impact of artificial light scatter on cup-to-disc area ratio, cup volume, rim volume, cup shape measure, height variation contour, and mean retinal nerve fiber layer (RNFL) thickness was evaluated.
Results:
Repeated-measures analysis of variance showed that there was no significant change in cup-to-disc area ratio, cup volume, rim volume, height variation contour, cup shape measure, or mean RNFL thickness as a function of increasing light scatter cell concentration.
Conclusion:
Artificial light scatter had no statistically significant impact on the stereometric parameters of the HRT II. From a clinical perspective, useful SLT data can be acquired with confidence from patients with diagnosed/suspected glaucoma and concomitant cataract.
More Related Videos
11:57Measuring Spatially- and Directionally-varying Light Scattering from Biological Material
Published on: May 20, 2013
10:42In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography (CT) and Light Microscopy (LM) Correlated with Scanning Electron Microscopy (SEM)
Published on: June 16, 2016
Related Concept Videos
Confocal Fluorescence Microscopy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...