Can aneroid sphygmomanometers be used at altitude?
N A Kametas1, F McAuliffe, E Krampl
1Harris Birthright Research Centre for Fetal Medicine, King's College Hospital, Denmark Hill, London, UK. n.kametas@btinternet.com
Mercury-free aneroid devices offer accurate blood pressure readings at high altitudes. This study validates a portable aneroid device, finding it reliable for clinical use in populations living at high elevations.
Area of Science:
- Clinical Medicine
- Biomedical Engineering
- Cardiovascular Research
Background:
- Growing concerns over mercury's environmental and health impacts are driving the adoption of mercury-free medical devices.
- Accurate blood pressure measurement is crucial, especially in diverse environmental conditions like high altitude.
Purpose of the Study:
- To evaluate the accuracy of a portable aneroid sphygmomanometer for blood pressure measurement in adults residing at high altitude.
- To assess device performance following British Hypertension Society protocol standards.
Main Methods:
- Calibration of the aneroid device at sea level and high altitude (4370 m).
- Validation using parallel comparisons with mercury sphygmomanometers in 10 subjects.
- Sequential and simultaneous same-arm measurements by two trained observers.
Main Results:
- All calibration comparisons between aneroid and mercury devices were within 3 mm Hg.
- The aneroid device achieved an 'A' grade for both systolic and diastolic pressures.
- Mean differences and standard deviations confirmed high accuracy in both sequential and simultaneous analyses.
Conclusions:
- The Riester-Exacta portable aneroid device is accurate and reliable for blood pressure measurement in adults at high altitude.
- This mercury-free device meets clinical validation standards, supporting its use in high-altitude environments.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: