Detection of refrigerator-associated 60 Hz alternating current as ventricular fibrillation by an implantable
Ayman S Al Khadra1, Abdulaziz Al Jutaily, Salem Al Shuhri
1Department of Adult Cardiology, Prince Sultan Cardiac Center, PO Box 7897, X-982, Riyadh 11159, Saudi Arabia. avnrt@hotmail.com
Abstract:
This report describes a patient with an implantable defibrillator who suffered an inappropriate defibrillation shock upon retrieving some food items from his inadequately earthed refrigerator. Noise typical of electrical interference can be observed in the stored electrogram of the episode. The patient was instructed to earth his home appliances, but he decided to avoid his refrigerator altogether, and has had no subsequent shocks.