Truncation artifact on PET/CT: impact on measurements of activity concentration and assessment of a correction
Osama Mawlawi1, Jeremy J Erasmus, Tinsu Pan
1Department of Imaging Physics, M. D. Anderson Cancer Center, 1515 Holcombe Blvd., Box 56, Houston, TX 77030, USA. omawlawi@mdanderson.org
Objective:
Discrepancy between fields of view (FOVs) in a PET/CT scanner causes a truncation artifact when imaging extends beyond the CT FOV. The purposes of this study were to evaluate the impact of this artifact on measurements of 18F-FDG activity concentrations and to assess a truncation correction algorithm.
Materials And Methods:
Two phantoms and five patients were used in this study. In the first phantom, three inserts (water, air, bone equivalent) were placed in a water-filled cylinder containing 18F-FDG. In the second phantom study, a chest phantom and a 2-L bottle fitted with a bone insert were used to simulate a patient's torso and arm. Both phantoms were imaged while positioned centrally (baseline) and at the edge of the CT FOV to induce truncation. PET images were reconstructed using attenuation maps from truncated and truncation-corrected CT images. Regions of interest (ROIs) drawn on the inserts, simulated arm, and background water of the baseline truncated and truncation-corrected PET images were compared. In addition, extremity malignancies of five patients truncated on CT images were reconstructed with and without correction and the maximum standard uptake values (SUVs) of the malignancies were compared.
Results:
Truncation artifact manifests as a rim of high activity concentration at the edge of the truncated CT image with an adjacent low-concentration region peripherally. The correction algorithm minimizes these effects. Phantom studies showed a maximum variation of -5.4% in the truncation-corrected background water image compared with the baseline image. Activity concentration in the water insert was 6.3% higher while that of air and bone inserts was similar to baseline. Extremity malignancies showed a consistent increase in the maximum SUV after truncation correction.
Conclusion:
Truncation affects measurements of 18F-FDG activity concentrations in PET/CT. A truncation-correction algorithm corrects truncation artifacts with small residual error.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...