Stability of the EasyOne ultrasonic spirometer for use in general practice.

Julia A E Walters1, Richard Wood-Baker, Justin Walls

  • 1Cardio-Respiratory Research Group, School of Medicine, University of Tasmania, Hobart, Tasmania, Australia.

Respirology (Carlton, Vic.)
|April 26, 2006
PubMed
Summary

This study found that the EasyOne portable spirometer maintains accuracy without daily calibration, supporting its use in general practice for diagnosing respiratory diseases.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same authorSame journal

Biomarkers of Pre-COPD Lung Function Trajectories in Middle Age: A Prospective Study From 1st to 6th Decade.

Respirology (Carlton, Vic.)·2026
Same author

Reply: An index of maximal expiratory flow-volume curve with different features related to central and peripheral concavity.

ERJ open research·2026
Same author

Concavity of the maximal expiratory flow-volume curve, and incidence of COPD and respiratory symptoms: a population-based cohort study.

ERJ open research·2026
Same author

Paradigm shifts in asthma: what is old is new again.

The Lancet. Respiratory medicine·2025
Same author

Paternal prepubertal passive smoke exposure is related to impaired lung function trajectories from childhood to middle age in their offspring.

Thorax·2025
Same author

Lifetime Body Mass Index Trajectories and Contrasting Lung Function Abnormalities in Mid-Adulthood: Data From the Tasmanian Longitudinal Health Study.

Respirology (Carlton, Vic.)·2025

Area of Science:

  • Pulmonary Medicine
  • Medical Device Technology

Background:

  • Spirometry is crucial for diagnosing and managing chronic respiratory diseases.
  • Current guidelines from the American Thoracic Society and European Respiratory Society (ATS/ERS) recommend daily spirometer calibration.
  • Many general practitioners do not meet this daily calibration recommendation.

Purpose of the Study:

  • To evaluate the accuracy and calibration stability of the portable ultrasonic EasyOne spirometer.
  • To assess if the EasyOne spirometer requires regular calibration as per ATS/ERS guidelines.

Main Methods:

  • Six EasyOne spirometers were tested in a practice-based setting.
  • Accuracy was assessed using a certified 3-L syringe for inspiratory and expiratory maneuvers.
  • Calibration checks were performed periodically over 26 weeks using dedicated and random mouthpiece assemblies.

Related Experiment Videos

Main Results:

  • All spirometers met ATS/ERS accuracy criteria throughout the study period.
  • No significant change in calibration was observed during 26 weeks of routine clinical use.
  • The spirometer's accuracy was unaffected by flow rates generated during calibration.

Conclusions:

  • The EasyOne spirometer demonstrated stable calibration during routine general practice use.
  • The findings support the manufacturer's claim that daily calibration is not required for this device.
  • This may simplify spirometry procedures in community settings.