Patient Alert to detect ICD lead failure: efficacy, limitations, and implications for future algorithms
Dirk Vollmann1, Ali Erdogan, Ewald Himmrich
1Abteilung Kardiologie und Pneumologie, Herzzentrum, Georg-August-Universität Göttingen, Germany. dirkvollmann2000@aol.com
Summary
Patient Alert detected one-third of implantable cardioverter-defibrillator (ICD) lead failures. Combining it with sensing integrity counter (SIC) monitoring improved detection to 100%, enhancing patient safety.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Implantable cardioverter-defibrillators (ICDs) are crucial for managing cardiac arrhythmias.
- Lead dysfunction is a significant complication affecting ICD performance and patient safety.
- Early detection of lead failures is essential for timely intervention and improved patient outcomes.
Purpose of the Study:
- To evaluate the effectiveness of the Patient Alert algorithm in detecting implantable cardioverter-defibrillator (ICD) lead failures.
- To assess the utility of Patient Alert for identifying lead dysfunction in a prospective cohort.
- To compare the performance of Patient Alert with continuous lead integrity monitoring (SIC).
Main Methods:
- A prospective study involving 360 ICD patients followed for 22 months.
- Patient Alert algorithm monitored lead impedance for abnormal values.
- Sensing Integrity Counter (SIC) was used for retrospective analysis of lead integrity.
- Performance metrics including sensitivity and positive predictive value (PPV) were calculated.
Main Results:
- Patient Alert detected 3 of 10 ICD lead failures.
- The Sensing Integrity Counter (SIC) identified an additional 4 lead defects retrospectively.
- Combined use of Patient Alert and SIC detected 100% of ICD lead failures.
- The combination also identified 71.4% of severe lead complications.
Conclusions:
- Patient Alert alone detects approximately one-third of ICD lead failures.
- Integrating SIC with Patient Alert significantly enhances sensitivity for lead failure detection.
- False positives with SIC may occur with integrated bipolar leads.
- Further improvements in Patient Alert could incorporate SIC and automated threshold measurements for better complication detection.