Matrix solid-phase dispersion microextraction and determination of pesticide residues in medicinal herbs by gas chromatography with a nitrogen-phosphorus detector
Related Experiment Video
Updated: Aug 9, 2026

An Efficient and Rapid HPLC-QQQ-MS Method for the Quantitation of Tropane Alkaloids in Medicinal Plants
Published on: June 12, 2026
Feng Tang1, Yongde Yue, Rimao Hua
1Anhui Agricultural University, College of Resources and Environment, Hefei, Anhui 230036, People's Republic of China.
Abstract:
A gas chromatography method for the determination of residues of 6 pesticides in Isatis indigotica Fort and Paeonia lactiflora Pall is described. Optimizations of different parameters were performed. The method used 0.5 g herb sample, silica gel as the matrix solid-phase dispersion sorbent, and acetone as the eluting solvent. Recoveries at spiked concentrations between 0.1 and 5 mg/kg were between 80.63 and 110.12%, with relative standard deviations ranging from 0.43 to 17.67%. The limits of quantitation were 0.01 mg/kg for metalaxyl, triadimefon, and paclobutrazol and 0.05 mg/kg for vinclozolin, tebuconazole, and fenarimol. The method may be useful as a screening protocol for the determination of these pesticides in medicinal herb samples.
More Related Videos
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Overview of Detectors