Blanked atrial flutter in patients with cardiac resynchronization therapy: clinical significance and implications for
Bernhard Strohmer1, Christiana Schernthaner, Maximilian Pichler
1Department of Cardiology, Salzburger Landeskliniken, Paracelsus Private Medical University, Salzburg, Austria. B.Strohmer@salk.at
Background:
Atrial arrhythmias are frequently observed in patients with heart failure and may be a primary cause for decompensation during cardiac resynchronization therapy (CRT). The accurate detection of organized atrial tachyarrhythmias poses a challenge to the function of mode-switching biventricular pacemakers/defibrillators.
Methods:
The purpose of the study was to determine retrospectively the incidence of blanked atrial flutter and mode switch failure (2:1 lock-in), and to look for factors predisposing to this problem. A total number of 65 patients with CRT devices has been followed regularly over 18 +/- 12 months. Five patients were excluded because of chronic atrial fibrillation and reprogramming to VVIR mode.
Results:
Seven out of 60 patients (12%) were diagnosed with blanked atrial flutter at unscheduled device interrogation. Sustained biventricular pacing at a median rate of 125/min-mimicking sinus tachycardia-resulted in rapid deterioration of heart failure and hospitalization. Mode switch failure occurred due to coincidence of every second flutter wave with atrial blanking. The group with 2:1 lock-in was programmed to longer atrial blanking times (143 +/- 34 ms vs 105 +/- 32 ms; P = 0.026) and AV intervals (126 +/- 8 ms vs 107 +/- 29; P = 0.001) than the group without lock-in. Other clinical characteristics examined did not differ between the two groups apart from a previous history of atrial fibrillation (P = 0.032).
Conclusion:
Blanked atrial flutter with rapid ventricular pacing is a clinically important problem in heart failure patients treated with CRT devices. Efforts should be made to avoid this complication by atrial lead implantation without ventricular farfield oversensing, by programming short PVAB and AV intervals, and by implementation of dedicated device algorithms.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
08:05Optimization of Transesophageal Atrial Pacing to Assess Atrial Fibrillation Susceptibility in Mice
Published on: June 29, 2022
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias VI: Management of Dysrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Mechanism of Cardiac Arrhythmias
Antiarrhythmic Drugs: Class II Agents as β-Adrenergic Blockers