Excitation of the intrinsic conduction system through his and interventricular septal pacing
Timothy G Laske1, Nicholas D Skadsberg, Alexander J Hill
1Medtronic, Inc., Minneapolis, Minnesota, USA.
Background:
Direct His bundle pacing results in rapid synchronous ventricular activation. However, clinical experiences with such pacing have been associated with long procedure times and compromised pacing and sensing performance.
Methods:
We evaluated myocardial activation sequences (AS) for pacing of the His bundle and peri-His region and assessed acute pacing performance using custom-designed plunge electrodes. Unipolar pacing was performed in isolated swine hearts (n = 10) using four quadripolar stimulation/sensing electrodes implanted into the interventricular septum and equally spaced between the membranous septum and the coronary sinus ostium (zones 1-4, respectively; electrode depth (ED) 1 = most distal, ED 4 = most proximal). Optimal pacing sites were defined as: pacing thresholds < or = 1.5 V, a P-R ratio of < or = 0.5, and > or = 50% occurrence of an intrinsic midseptal left ventricular (LV) endocardial electrical breakout (BO) and activation pattern.
Results:
Pacing thresholds improved with greater depth of electrode location within the septum (ED 1: 1.51 +/- 0.8 V vs ED 4: 5.2 +/- 3.8 V, P < 0.001), as did the P-R ratio (0.34 +/- 0.6 vs 0.78 +/- 1.0, P < 0.05). His potentials were only observed in zone 1 and 2 electrodes (0.12 and 0.02 mV, respectively). Only electrodes in zones 1 and 2 produced LV endocardial electrical BOs in the midseptal region that demonstrated an intrinsic-like endocardial AS. Depth 1 and 2 electrodes (11.75 and 8.75 mm, respectively) in zone 1 satisfied all three optimal pacing site requirements.
Conclusions:
This study has shown that LV activation patterns similar to sinus rhythm may be achieved without direct activation of the His bundle, while maintaining acceptable pacing and sensing performance. These data indicate that pacing systems designed to stimulate the tissues below the point at which the His bundle penetrates the central fibrous body may provide improved system efficiency and LV performance in comparison to both direct His bundle pacing and traditional pacing sites.
Related Concept Videos
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
The Cardiac Cycle
The Process
Electrical signals—sent from the sinoatrial (SA) node in the right atrial wall to the atrioventricular (AV) node between the right atrium and right ventricle—cause both atria to simultaneously contract. When the signal reaches the AV node, it pauses for approximately a tenth of a second, allowing the atria to contract and empty blood into the...
Electrophysiology of Normal Cardiac Rhythm
Mechanism of Cardiac Arrhythmias
Cardiac Action Potential
The cardiac action potential process involves a series of phases characterized by the movement of ions across the cardiac cell membranes, leading to the depolarization and repolarization of the cardiac myocytes.
Ionic Basis of Cardiac Action Potentials