Pacemaker tachycardia in a minute ventilation rate-adaptive pacemaker induced by electrocardiographic monitoring
Wendy Lau1, Susan J Corcoran, Harry G Mond
1Department of Cardiology, The Royal Melbourne Hospital, Victoria 3050, Australia.
Abstract:
Programmed upper rate pacing occurred in a patient with a rate-adaptive pacemaker when he was connected to a cardiac monitor in an emergency department. The tachycardia was mistakenly interpreted to be ventricular tachycardia and the patient received multiple DC shocks as well as intravenous amiodarone and sotalol, resulting in severe hemodynamic deterioration. It is important that physicians working in the hospital environment be familiar with this pacemaker-monitor interaction as the problem may be easily rectified by disconnecting the monitor or by reprogramming the pacemaker to a nonrate-adaptive pacing mode.
Related Concept Videos
Dysrhythmias III: Characteristics of Dysrhythmias
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Respiratory Volumes and Capacities I
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias