Successful thrombus extraction with the Rescue thrombus management system during acute percutaneous coronary

Kunadian Vijayalakshmi1, Babu Kunadian, Robert A Wright

  • 1Department of Cardiology, the James Cook University Hospital, Middlesbrough, United Kingdom.

Insights

The Rescue device effectively removed debris and improved coronary blood flow in most patients undergoing percutaneous coronary intervention for acute coronary syndrome. Further research is needed to confirm its impact on tissue perfusion and clinical outcomes.

Area of Science:

  • Cardiology
  • Interventional Cardiology
  • Medical Devices

Background:

  • Percutaneous coronary intervention (PCI) is a common treatment for acute coronary syndrome.
  • Thrombus-laden coronary arteries pose challenges during PCI, potentially leading to suboptimal outcomes.
  • Effective debris removal is crucial for successful PCI and improved myocardial perfusion.

Purpose of the Study:

  • To evaluate the effectiveness of the Rescue device in restoring coronary blood flow and achieving optimal tissue perfusion.
  • To assess the device's ability to aspirate debris during PCI in patients with thrombus-laden coronary arteries.

Main Methods:

  • A prospective study involving 30 patients with acute coronary syndrome undergoing PCI.
  • Utilized the Rescue device for potential thrombus aspiration prior to or during PCI.
  • Assessed angiographic outcomes including TIMI frame count, thrombus score, TIMI flow grade, and myocardial blush grade.

Main Results:

  • The Rescue device successfully aspirated debris in 87% of cases.
  • Significant improvements were observed in TIMI frame count (P < 0.0001) and thrombus score (P < 0.0001) post-Rescue device use.
  • TIMI flow grade 3 was achieved in 60% post-Rescue and 87% post-PCI; myocardial blush grade 3 in 13% post-Rescue and 37% post-PCI.

Conclusions:

  • The Rescue device demonstrates capability in aspirating significant amounts of debris.
  • The device effectively improves angiographic epicardial coronary blood flow.
  • The clinical impact of the Rescue device on tissue perfusion and patient outcomes requires further investigation.
Abstract