Related Experiment Video
Updated: Aug 8, 2026

Ultrasound Velocity Measurement in a Liquid Metal Electrode
Published on: August 5, 2015
New flowmetric measurement methods of power dissipated by an ultrasonic generator in an aqueous medium
Valérie Mancier1, Didier Leclercq
1D.T.I., E.A. 3083, UFR Sciences Exactes et Naturelles, BP 1039, F-51687 Reims Cedex 2, France.
Abstract:
Two new determination methods of the power dissipated in an aqueous medium by an ultrasound generator were developed. They are based on the use of a heat flow sensor inserted between a tank and a heat sink that allows to measure the power directly coming through the sensor. To be exploitable, the first method requires waiting for stationary flow. On the other hand, the second, extrapolated from the first one, makes it possible to determine the dissipated power in only five minutes. Finally, the results obtained with the flowmetric method are compared to the classical calorimetric ones.