Epicardial radiofrequency ablation of ventricular myocardium: factors affecting lesion formation and damage to
Guilherme Fenelon1, Kleber Ponzi Pereira, Angelo A V de Paola
1Department of Cardiology, Paulista School of Medicine, Federal University of São Paulo, São Paulo, SP, Brazil. guilhermefenelon@uol.com.br
Unlabelled:
We evaluated the factors affecting epicardial radiofrequency (RF) lesion formation in normal ventricular myocardium. In 16 dogs, a minithoracotomy was made and a sheath was placed in the pericardial space. Standard ablation lesions (4-mm tip catheter; 70 ( composite function) C/60 seconds) were created in each ventricle under fluoroscopy guidance (n = 7) or hand-held with direct visualization of the catheter to assure optimal electrode-tissue contact (n = 6). In the latter, thermally-shielded (TS) electrodes (50% tip surface along its 4 mm length) were used in 3/6 dogs. Catheter tip (4 mm) irrigation (13 mL/minutes; 40 ( composite function) C/60 seconds) was employed with conventional techniques in 3 additional dogs.
Results:
With optimal electrode-tissue contact (11 lesions), power (3.4 +/- 2.3 W vs. 16 +/- 13 W; p < 0.001) and pacing thresholds (0.2 +/- 0.0 mA vs. 3.6 +/- 5.7 mA; p = 0.004) were lower than standard RF (25 lesions). However, lesion dimensions were similar and transmural lesions did not occur (depth 2.8 +/- 1.1 mm vs. 3.0 +/- 1.5 mm). Catheter irrigation allowed high power outputs (43 +/- 6.1 W; p < 0.001) generating transmural lesions, 5/9 (55%), depth 6.4 +/- 2.1 mm. At constant power (2 W), catheter-tip temperature (52 +/- 5.2( composite function) C vs. 57 +/- 6.6( composite function) C; p = NS) and lesion (10 in each group) dimensions were similar for conventional and TS electrodes, but damage to parietal pericardium and lungs occurred with conventional electrodes only (70% vs. 0% p = 0.02).
Conclusion:
Standard epicardial RF ablation does not produce deep lesions and exhibits a significant energy loss probably due to poor electrode-tissue contact. Catheter irrigation allows delivery of high power outputs to the epicardium consistently creating deeper lesions than standard ablation. TS electrodes may reduce damage to neighboring structures during epicardial RF ablation.
More Related Videos
05:43Determination of Continuity Index Values in Atrial Fibrillation Ablation with Proactive Esophageal Cooling
Published on: April 19, 2024
06:25Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Related Concept Videos
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Layers of the Heart Wall
The myocardium, the thickest layer, consists of cardiac muscle cells interconnected by intercalated discs and crisscrossing connective tissue fibers. These muscle fibers contract...