Related Experiment Video
Updated: Aug 8, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
High failure rate of a modern, proximally roughened, cemented stem for total hip arthroplasty
A Grose1, A González Della Valle, P Bullough
1Hospital for Special Surgery, New York, NY 10021, USA.
Abstract:
The role of surface finish on the survivorship of cemented femoral stems continues to be debated. A total of 34 proximally roughened cemented stems were implanted in 33 consecutive patients undergoing total hip arthroplasty by a single surgeon. An alarmingly high failure rate was observed, prompting a retrospective chart review, analysis of radiographs, and evaluation of retrieved stems and pathological specimens. Nineteen patients were available with more than two years follow-up. Of these 19 patients, nine stems had failed (47%) due to severe osteolysis and stem loosening. Failures were significantly more common in the male gender (p<0.005), and young (p=0.05), tall (p<0.002), and heavy patients (p<0.004). All failed revised hips showed severe metallosis, with both gross and microscopic evidence of metallic shedding from the stems. Our findings suggest that this proximally roughened stem is susceptible to early failure. Failure is characterized by stem debonding, subsidence within the cement mantle, shedding of metallic and cement particles due to fretting, and rapidly progressive osteolysis. These findings have been observed with other rough surface finish cemented stems.