Related Experiment Video
Updated: Aug 8, 2026

Subject-specific Musculoskeletal Model for Studying Bone Strain During Dynamic Motion
Published on: April 11, 2018
Ligament strains predict knee motion after total joint replacement: a kinematic analysis of the sigma knee
Elvis C S Chen1, Joel L Lanovaz, Randy E Ellis
1School of Computing, Queen's University, Kingston, Ontario, Canada. chene@cs.queensu.ca
Abstract:
A passive forward kinematics knee model was used to predict knee motion of a total joint replacement. Given ajoint angle, maps of articular surfaces, and patient-specific ligament properties, this model predicted femorotibial contact locations based on the principle of ligament-strain minimization. The model was validated by physical experiments on a commonly implanted knee prosthesis, showing excellent correspondence between the model and actual physical motion. Results suggest that the knee prosthesis studied required an intact posterior cruciate ligament to induce the desirable roll-back motion, and that a single-bundle model of major knee ligaments generated kinematics similar to that of a multibundle ligament model. Implications are that a passive model may predict knee kinematics of a given patient, so it may be possible to optimize the implantation of a prosthesis intraoperatively.