Related Experiment Video
Updated: Aug 8, 2026

Oxygenation-sensitive Cardiac MRI with Vasoactive Breathing Maneuvers for the Non-invasive Assessment of Coronary Microvascular Dysfunction
Published on: August 17, 2022
Inter-breath-hold registration for the production of high resolution cardiac MR volumes
Nicholas M I Noble1, Redha Boubertakh, Reza S Razavi
1Centre for Medical Image Computing, University College London, UK. N.Noble@ucl.ac.uk
Abstract:
High resolution MRI images of the beating heart permit observation of detailed anatomical features and enable quantification of small changes in metrics of cardiac function. To obtain approximately isotropic sampling with an adequate spatial and temporal resolution, these images need to be acquired in multiple breath-holds. They are, therefore, often affected by through-plane discontinuities due to inconsistent breath-hold positions. This paper presents a method to correct for these discontinuities by performing breath-hold-by-breath-hold registration of high resolution 3D data to radial long axis images. The corrected images appear free of discontinuities, and it was found that they could be delineated more reproducibly than uncorrected images. This reduces the sample size required to detect systematic changes in blood pool volume by 57% at end systole and 78% at end diastole.