Radiochromic films for improved evaluation of patient dose in liver interventions
Laura Mantovani1, Loredana D'Ercole, Francesco Lisciandro
1Departments of Medical Physics, University of Pavia, Italy.
Purpose:
To investigate a method for evaluation of the maximum skin dose (MSD) of radiation in patients undergoing interventional radiology procedures in the liver.
Materials And Methods:
Transarterial chemoembolization, transarterial embolization, portal vein embolization, and biliary interventions were the procedures considered in this study. Ninety procedures in 70 patients were studied. The complexity of the biliary interventions was taken into account during the analysis. The MSD was measured with use of GAF chromic XR type R films, whereas the dose-area product (DAP) was measured by a transmission chamber.
Results:
The MSD was measured in a group of 19 patients. The coefficient of the interpolation line between the skin dose and the DAP (0.0051 cm(-2)) was determined. An approximated value of MSD from the DAP for the remaining 71 procedures was estimated by means of an interpolation line. The mean MSD in the endovascular procedures was 4.4 Gy; for the three different levels of complexity of the biliary procedures, the mean MSDs were 0.4, 1.2, and 3 Gy, respectively.
Conclusions:
Radiochromic films are an easy-to-use and efficient method for measuring skin entrance radiation dose and have the advantage of providing information on the MSD as well as the distribution of radiation to the skin. In light of these results, the authors suggest recording the dose for biliary and endovascular procedures as indicated by the International Commission on Radiological Protection and the United States Food and Drug Administration.
More Related Videos
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...