A study of the effect of setup errors and organ motion on prostate cancer treatment with IMRT
Valeria Landoni1, Bianca Saracino, Simona Marzi
1Laboratorio di Fisica Medica e Sistemi Esperti, Istit4uto Regina Elena, Rome, Italy. landoni@ifo.it
Purpose:
To assess the influence of setup errors and organ motion in terms of the probability of tumor control and normal-tissue complications by tumor control probability and normal-tissue complication probability.
Methods And Materials:
Twelve patients were treated for prostate cancer with intensity-modulated radiation therapy. Two orthogonal portal images were taken daily. All patients underwent three computed tomography scans during the 8-week treatment time (i.e., baseline, intermediate, and final). The original treatment plans were re-evaluated, taking into account setup errors and organ motion.
Results:
The mean shifts +/- standard deviation of the whole patient population in the lateral, anterior-posterior, and craniocaudal direction were 1.0 +/- 1.5 mm, 0.9 +/- 2.1 mm, and 1.9 +/- 2.1 mm, respectively. In most of the recalculated dose-volume histograms, the coverage of clinical target volume was granted despite organ motion, whereas the rectal wall histograms were often very different from the planned ones.
Conclusion:
We have studied the impact of prostate and rectum motion, as well as setup errors, on dose-volume histograms. The estimate of these effects may have implications for predictive indications when planning intensity-modulated radiation therapy treatments on prostate.