A novel approach to accurate portal dosimetry using CCD-camera based EPIDs
E M Franken1, J C J de Boer, B J M Heijmen
1Department of Radiotherapy, Division of Clinical Physics, Erasmus MC-Daniel den Hoed Cancer Center, Groene Hilledijk 301, 3075 EA Rotterdam, The Netherlands. e.franken@erasmusmc.nl
This study introduces a novel portal dosimetry method using cooled CCD camera-based electronic portal imaging devices (CEPIDs). It accurately calculates portal dose images (PDIs) without prior assumptions, enabling faster and more precise radiotherapy quality assurance.
Area of Science:
- Medical Physics
- Radiotherapy Technology
- Image Processing
Background:
- Accurate portal dosimetry is crucial for radiotherapy quality assurance.
- Existing methods for electronic portal imaging devices (EPIDs) often rely on a priori assumptions about cross-talk.
- CCD camera-based EPIDs (CEPIDs) offer high signal-to-noise ratios for improved imaging.
Purpose of the Study:
- To develop and validate a new portal dosimetry method using CEPIDs.
- To determine and correct for non-symmetrical and position-dependent cross-talk in CEPIDs.
- To enable fast and accurate derivation of portal dose images (PDIs) from electronic portal images (EPIs).
Main Methods:
- Directly imaged cross-talk kernels from small fields ('pencil beams') using a cooled CCD camera.
- Signal calibration using two reference fields.
- Iterative deconvolution to derive PDIs from EPIs, accounting for determined cross-talk.
Main Results:
- The method achieved high accuracy (0.2+/-0.7% 1 SD) across various field configurations and phantom types.
- Validated against scanning diode measurements for diverse clinical scenarios, including IMRT fields.
- Identified back-scattered radiation as a significant contributor to CEPID cross-talk signals.
Conclusions:
- The proposed method provides a fast and accurate approach to portal dosimetry using CEPIDs.
- It overcomes limitations of previous methods by directly characterizing cross-talk.
- Offers a robust solution for real-time quality assurance in radiotherapy.