Pulse transit time based on piezoelectric technique at the radial artery
Jong Yong Abdiel Foo1, Chu Sing Lim
1Biomedical Engineering Research Centre, Nanyang Technological University, 50 Nanyang Drive, Research Techno Plaza, 6th Storey, XFrontiers Block, Singapore, 637553, Singapore. jong@ntu.edu.sg
Objectives:
Pulse transit time (PTT) has shown its potential in relevant cardiovascular and cardiorespiratory studies. However, the use of photoplethysmography (PPG) in PTT measurement can be limited in events of poor peripheral perfusion. Uninterrupted PTT monitoring may also not be achievable when less cooperative patients distribute the PPG probe due to its prominent light source. Hence, there is a need for an alternative method to measure PTT in such incidents.
Methods:
In this study, the piezoelectric (PIEZO) technique to detect pulsations from a human wrist above the radial artery to estimate PTT is presented. 17 healthy adults (11 male; age range of 21-33 years) were recruited to compare PTT and heart rate (HR) differences between the PPG and PIEZO methods. These time-related derivations were made with respect to an electrocardiogram (ECG).
Results:
The timing consistency of the PIEZO transducer shows significant correlations (p < 0.01) to those derived from the ECG and a pulse oximeter. Particularly, there is a high level of agreement of < 1 beat per minute (bpm) difference in HR estimates observed when compared to the two commercial devices in the respective Bland-Altman plots. Comparison of PTT obtained from the PIEZO transducer against the PPG signal shows constantly lower values due to the shorter path length it requires to propagate. A regression equation was formulated to relate the PTT values acquired from both these signals.
Conclusions:
Preliminary findings herein suggest that the PIEZO technique can be useful as an alternative for PTT monitoring. This shows promise to be more accommodating for less cooperative patients or those with insufficient peripheral perfusion.
Related Concept Videos
Pulse
The pulse serves as a clinical indicator...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Pulse Assessment Sites
Special considerations while measuring pulse