Related Experiment Video
Updated: Aug 8, 2026

Electrophysiological Methods to Assess Peripheral Pain Block in an Anesthetized Rat
Published on: November 21, 2025
An objective assessment of nerve stimulators used for peripheral nerve blockade*
D Jochum1, G Iohom, D P Diarra
1Groupe Hospitalier Privé du Centre Alsace, Colmar, France. jochumd@calixo.net
Abstract:
Nerve stimulation is considered by many to be the current 'gold standard' for locating peripheral nerves. Thirteen nerve stimulators were subjected to a battery of tests performed by two independent observers under standardised conditions using a digital oscilloscope, a calibrated resistance and a novel scoring system. Individual scores were assigned for the signal waveform, current intensity, impulse duration, maximum load output and functionality of each nerve stimulator; the maximum score achievable was 20. A score of > or = 15/20 was achieved by six nerve stimulators: Stimuplex HNS 12 (B Braun); MultiStim Vario (Pajunk); Plexival (Medival, Vygon); MultiStim Sensor (Pajunk); Plexygon (Aryon, Vygon); Stimuplex HNS 11 (B Braun). Seven nerve stimulators scored < 15/20. Anaesthetists should be aware of the limitations of the nerve stimulator being used. Standardisation of features is desirable when manufacturing nerve stimulators.