Simulated clinical evaluation of four fluid warming devices*
M Turner1, I Hodzovic, W W Mapleson
1Anaesthesia and Intensive Care, Royal Gwent Hospital, Newport, UK. matthewwturner@hotmail.com
Abstract:
The fluid warming capabilities of the Bair Hugger, Hotline, Standard Ranger and Fluido devices were evaluated in the laboratory with gravity flow via a 14G cannula (1 m head of fluid) and with the fluid bag pressurised to 300 mmHg. The resulting flows (70-450 ml.min(-1)) were recorded. At a room temperature of 22 degrees C, no device warmed the fluid to 37 degrees C. However, the Standard Ranger with gravity flow and the Fluido with both gravity and pressurised flow achieved 35 degrees C, whereas the Bair Hugger and Hotline with both gravity and pressurised flow, and the Standard Ranger with pressurised flow, achieved only 24-31 degrees C. However, from the way delivered temperatures changed with flow, we calculated that all four devices would achieve close to 37 degrees C at the flows specified by the manufacturers: 17, 83, 150 and 800 ml.min(-1) for the Bair Hugger, Hotline, Standard Ranger and Fluido, respectively.
Related Concept Videos
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...