Effect of electrode configuration on psychophysical forward masking in cochlear implant listeners
Bom Jun Kwon1, Chris van den Honert
1Cochlear Americas, 400 Inverness Parkway, Suite 400, Englewood, Colorado 80112, USA. bjkwon@gmail.com
Abstract:
Bipolar stimulation has been thought to be more beneficial than monopolar stimulation for speech coding in cochlear implants, on the basis of its more restricted current flow. The present study examined whether bipolar stimulation would indeed lead to reduced channel interaction in a behavioral forward masking experiment tested in four Nucleus 24 users. The masker was fixed on one channel and three masker levels that were balanced for loudness between the configurations were chosen. As expected, masking was maximal when the masker and probe channels were spatially close and decreased as they were separated. However, overall masking patterns did not consistently demonstrate sharper tuning with bipolar stimulation than monopolar. This implies that the spatial extent of a bipolar current field is not consistently narrower than that of an equally loud monopolar stimulus; therefore, it should not be assumed that bipolar stimulation leads to reduced channel interaction. Notably, bipolar masking patterns appeared to display more variations across channels, possibly influenced more by anatomical and neural irregularities near electrode contacts than monopolar masking patterns. The present psychophysical results provide a theoretical basis regarding the widespread use (and success) of monopolar configurations by implant users.