PET/CT Biograph Sensation 16. Performance improvement using faster electronics.

M-J Martínez1, Y Bercier, M Schwaiger

  • 1Nuclear Medicine Department, Klinikum rechts der Isar der Technischen Universitaet Muenchen, Ismaninger Strasse 22, 81675 Muenchen, Germany. m.j.martinez@lrz.tum.de

Nuklearmedizin. Nuclear Medicine
|May 20, 2006
PubMed
Summary

The Pico-3D electronics upgrade significantly improved PET scanner performance, enhancing image quality and count-rate capabilities. This upgrade allows for reduced scan times and increased patient throughput in PET imaging.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Prediction of synchronous distant metastasis of primary pancreatic ductal adenocarcinoma using the radiomics features derived from <sup>18</sup>F-FDG PET and MRI.

Clinical radiology·2023
Same author

Stem Cell Therapy after Acute Myocardial Infarction-The Emperor's New Clothes?: Stem Cell Mobilization by Granulocyte Colony-Stimulating Factor in Patients with Acute Myocardial Infarction: A Randomized Controlled Trial. JAMA 295: 1003-1010, 2006.

Journal of the American Society of Nephrology : JASN·2023
Same author

Reply to "Antibiotics in orthognathic surgery and postoperative infections".

International journal of oral and maxillofacial surgery·2022
Same author

Surgical site infections in orthognathic surgery: prolonged versus single-dose antibiotic prophylaxis.

International journal of oral and maxillofacial surgery·2022
Same author

Determination of blood loss in bimaxillary surgery: does the formula and the time point affect results?

International journal of oral and maxillofacial surgery·2021
Same author

The sodium iodide symporter (NIS): novel applications for radionuclide imaging and treatment.

Endocrine-related cancer·2021

Area of Science:

  • Medical Imaging
  • Nuclear Medicine
  • Instrumentation

Background:

  • The Biograph Sensation 16 (BS16) PET/CT tomographs utilize detector electronics that influence timing coincidence windows and energy thresholds.
  • Previous performance benchmarks of the BS16 system are established.
  • Advancements in detector electronics aim to optimize PET scanner performance.

Purpose of the Study:

  • To evaluate the performance of the BS16 PET scanner before and after the Pico-3D electronics upgrade.
  • To quantify changes in key performance metrics such as scatter fraction, count-rate performance, and image quality.
  • To assess the impact of optimized signal processing on PET imaging protocols.

Main Methods:

  • Four NEMA NU 2-2001 protocols were employed: spatial resolution, scatter fraction, count losses and randoms, sensitivity, and image quality.

Related Experiment Videos

  • Performance evaluations were conducted on the BS16 PET scanner prior to and following the Pico-3D electronics installation.
  • Image reconstruction was performed using the AW-OSEM (4ix8s) algorithm.
  • Main Results:

    • A 14% decrease in scatter fraction was observed post-upgrade, with a value of approximately 0.42 at typical patient background activity.
    • Noise Equivalent Count-Rate (NECR) performance improved, with the peak R(NEC) increasing from 3.7 x 10(4) s(-1) to 6.4 x 10(4) s(-1).
    • Peak true count-rate increased from 1.9 x 10(5) s(-1) to 3.4 x 10(5) s(-1), and contrast improved by 45% for hot spheres and 12% for cold spheres.

    Conclusions:

    • The Pico-3D electronics upgrade significantly enhances BS16 PET scanner performance through optimized signal processing.
    • Narrower energy and timing coincidence windows result in a substantial increase in the signal-to-noise ratio.
    • The upgraded BS16 tomographs support imaging protocols with reduced acquisition times, leading to higher patient throughput.