Abstract:
A solid phase extraction method was developed for the preconcentration and separation of trace amounts of chromium, manganese, iron, cobalt, copper, cadmium and lead from environmental samples by complexation with alpha-benzoin oxime followed by adsorption onto Diaion SP-850-solid phase extraction column. One molar per liter HNO(3) was used as eluent. The recoveries of analytes at pH 8.0 with 700 mg of resin were greater than 95% without interference from alkaline, earth alkaline and some metal ions. The detection limits by three sigma for analyte ions were 0.65 microg l(-1) for Cr(III), 0.42 microg l(-1) for Mn(II), 0.28 microg l(-1) for Fe(III), 0.73 microg l(-1) for Co(II), 0.30 microg l(-1) for Cu(II), 0.47 microg l(-1) for Cd(II) and 0.50 microg l(-1) for Pb(II). The validation of the procedure was performed by the analysis of the certified standard reference materials. The presented procedure was applied to the determination of analytes in tap, river and sea waters, rice, wheat, canned tomato and coal samples with successfully results (recoveries greater than 95%, R.S.D.'s lower than 8%).
More Related Videos
10:44Clean Sampling and Analysis of River and Estuarine Waters for Trace Metal Studies
Published on: July 1, 2016
10:31Detection and Recovery of Palladium, Gold and Cobalt Metals from the Urban Mine Using Novel Sensors/Adsorbents Designated with Nanoscale Wagon-wheel-shaped Pores
Published on: December 6, 2015
Related Concept Videos
Extraction: Advanced Methods
Ion-Exchange Chromatography
Precipitation and Co-precipitation
Electrodeposition
Electrodeposition can...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Ion Exchange