Related Experiment Video
Updated: Aug 8, 2026

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Novel display technique for reference images for visibility of temporal change on radiographs--color digital
Yuji Ogata1, Hiroaki Naito, Hiroki Azuma
1Department of Medical Physics, Osaka University Graduate School of Medicine, Japan.
Purpose:
The present study investigated color digital summation radiography (CDSR) as a novel display technique for reference images for the visibility of temporal change on radiographs. In CDSR, only the parts with temporal differences are displayed in color. Moreover, all other parts are displayed in gray scale.
Materials And Methods:
CDSR was defined as "radiographs combined by the additive color mixture method". The visibility of simulated nodules located in the lung fields and mediastinum was evaluated by 12 radiologists (mean experience, 9.8 years; range, 1-26 years) for 24 conventional radiographs and CDSR for 6 color patterns. A five-point rating system (5, very good; 4, good; 3, adequate; 2, poor; 1, very poor) was used.
Results:
The mean scores (average +/- standard deviation) for the visibility of simulated nodules were as follows: magenta, 3.88 +/- 0.90; blue, 3.08 +/- 0.72; green, 3.04 +/- 0.86; red, 3.00 +/- 0.98; cyan, 2.71 +/- 0.86; and yellow, 2.50 +/- 0.72. Compared to conventional radiography in gray scale, at 1.21 +/- 0.41, all six color patterns for CDSR displayed significantly improved scores (p<0.001).
Conclusion:
CDSR might represent a useful technique for reference images from chest digital radiography.
More Related Videos
08:02Novel Quantification Protocol for Cardiovascular Calcification Progression Using Longitudinal MicroPET/MicroCT Images
Published on: November 15, 2024
05:49Reliability of Artificial Intelligence-Based Cone Beam Computed Tomography Integration with Digital Dental Images
Published on: February 23, 2024
Related Concept Videos
X-ray Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...