Related Experiment Videos
Failure to recognize loss of incoming data in an anesthesia record-keeping system may have increased medical
Michael M Vigoda1, David A Lubarsky
1Department of Anesthesiology, Perioperative Medicine and Pain Management, Center for Informatics and Perioperative Management, University of Miami/Jackson Memorial Medical Center, Miami, Florida 33136, USA. Mvigoda@med.miami.edu
Abstract:
Automated anesthesia record-keeping systems (AARKs) are increasingly being used. There is a perception that AARKs may limit medical liability. We report a case in which our AARK may have increased our medical liability exposure. Nine months after a patient suffered a serious intraoperative complication, the anesthesiologist was named (as one of several defendants) in a claim alleging failure to properly monitor anesthetic care. One reason why the anesthesiologist was named related to a gap of 93 min in which no vital signs were documented in the anesthesia record. Relying on the physiological monitors to assess the patient's condition, the anesthesiologist did not recognize the interruption of data transmission, because the "active" medication window obscured the graphical display of the vital sign window.
Related Concept Videos
Guidelines and Strategies for Safe Computer Charting
Maintain Confidentiality and Security:
Documentation of Nursing Diagnosis
In some settings, data-driven computerized decision support systems are in place, allowing for more accurate nursing diagnoses. The database within one of these systems includes diagnostic labels defining characteristics, activities, and indicators for nursing. A nurse enters assessment...
Legal Guidelines for Documentation
Data Reporting and Recording
Errors occurring during blood pressure monitoring