Related Experiment Video
Updated: Aug 8, 2026

Quantifying Inferior Vena Cava Compliance and Distensibility in an In Vivo Ovine Model Using 3D Angiography
Published on: April 26, 2024
Variations of patient doses in interventional examinations at different angiographic units
Doğan Bor1, Türkay Toklu, Turan Olğar
1Department of Engineering Physics, Ankara University Faculty of Engineering, 06100 Tandoğan, Ankara, Turkey. bor@eng.ankara.edu.tr
Purpose:
We analyzed doses for various angiographic procedures using different X-ray systems in order to assess dose variations.
Methods:
Dose-area product (DAP), skin doses from thermoluminescent dosimeters and air kerma measurements of 308 patients (239 diagnostic and 69 interventional) were assessed for five different angiographic units. All fluoroscopic and radiographic exposure parameters were recorded online for single and multiprojection studies. Radiation outputs of each X-ray system were also measured for all the modes of exposure using standard protocols for such measurements.
Results:
In general, the complexity of the angiographic procedure was found to be the most important reason for high radiation doses. Skill of the radiologist, management of the exposure parameters and calibration of the system are the other factors to be considered. Lateral cerebral interventional studies carry the highest risk for deterministic effects on the lens of the eye. Effective doses were calculated from DAP measurements and maximum fatal cancer risk factors were found for carotid studies.
Conclusions:
Interventional radiologists should measure patient doses for their examinations. If there is a lack of necessary instrumentation for this purpose, then published dose reports should be used in order to predict the dose levels from some of the exposure parameters. Patient dose information should include not only the measured quantity but also the measured radiation output of the X-ray unit and exposure parameters used during radiographic and fluoroscopic exposures.
Related Concept Videos
Dosage Regimen: Individualization
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System V: CT
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...