Related Experiment Video
Updated: Aug 8, 2026

Esophageal Heat Transfer for Patient Temperature Control and Targeted Temperature Management
Published on: November 21, 2017
Convection heating in pediatric general surgery - a comparison of warming alternatives in a mannequin study
Scott Nightingale1, Lisa Wynne, John Cassey
1Department of Paediatrics, John Hunter Hospital, NSW, Australia.
Background:
Numerous methods of patient warming are used to prevent intraoperative hypothermia in children. Commercially available forced air warming blankets are effective, but are single-use items. We tested a custom-designed heat dissipation unit (HDU) against one such commercially available blanket.
Methods:
Air temperatures at various points around a mannequin under simulated operating conditions were recorded using thermistors and thermal imaging. The only variable changed was the heating method: a forced air blanket or a customized HDU with two draping techniques - cotton drapes with and without a plastic 'undersheet'.
Results:
The three methods produced similar temperature increases and plateaux across the 11 thermistor points measured. There were no significant differences between temperatures at 1 h. A plastic sheet did not appear to enhance the effectiveness of the HDU in this study. Thermal imaging photography suggested more uniform heating of the mannequin with the HDU arrangements.
Conclusions:
The custom-built HDU compares favorably in our mannequin study with a Bair Hugger forced air warming blanket. As it is reusable, it offers considerable potential savings.
More Related Videos
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
10:23A Computational Modeling Approach to Investigate the Influence of Hyperthermia on the Tumor Microenvironment
Published on: December 1, 2023
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Decreased Body Temperature