Measuring the Carotid to Femoral Pulse Wave Velocity (Cf-PWV) to Evaluate Arterial Stiffness
Published on: May 3, 2018
Oscillometric blood pressure measurement used for calibration of the arterial tonometry method contributes
1Institute of Neurological Sciences, Southern General Hospital, Department of Clinical Physics, Glasgow, UK. s.hansen@clinmed.gla.ac.uk
Background And Objective:
We have investigated the accuracy of the Colin tonometry module, a non-invasive alternative to intra-arterial blood pressure (BP) measurement. The claims of previous publications range between good agreement to significant discrepancies between the Colin tonometry and intra-arterial methods. The Colin method relies on calibration of the tonometry module using an oscillometric method. Our aim was to look at the effect of this calibration on the accuracy.
Methods:
Ten conscious patients who already had an indwelling arterial catheter were studied. The Colin monitor was set to perform ten calibration cycles first with the oscillometric cuff on the same arm as the tonometry module and secondly on the opposite arm. Simultaneous recording of BP wave forms from the intra-arterial BP transducer and the Colin monitor allowed comparison of beat-by-beat systolic, diastolic and mean BP.
Results:
There was considerable inter- and intra-patient variation. In the worst case, the error range was 41 mmHg over ten calibrations. Analysis of variance showed that contra- and ipsilateral calibrations gave a significantly different bias while the multiple calibrations accounted for a significant proportion of the variability in systolic BP error.
Conclusions:
The Colin Tonometry method is not accurate enough to be used with confidence in clinical practice. The main reason for this is its reliance on an oscillometric method for calibration of the tonometry module. Single BP measurements, using either manual or semiautomatic instruments may vary considerably from the 'true' BP due to short-term perturbations of BP.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement