Factors that influence radiation dose in percutaneous coronary intervention
Virginia Tsapaki1, Athanassios Magginas, Eliseo Vano
1Department of Medical Physics, Konstantopoulio Agia Olga Hospital, Athens, Greece. virginia@otenet.gr
Aim:
To explore the factors that may influence the radiation dose imparted to the patient in PCI, and investigate whether the use of the latest digital X-ray system based on FP detector technology can have an impact on dose.
Materials And Method:
Demographic and clinical data such as number of lesions treated, number of stents placed, grade of tortuosity, and stage of occlusion, as well as use of double wire and double balloon technique, ostial stenting or bifurcation stenting, and presence of major complications were recorded, together with radiation parameters.
Results:
The factors that increased patient radiation dose were (1) patient gender, as men exhibited higher doses than women; (2) complex lesion; (3) increasing number of stents; (4) position of stent; (5) grade of tortuosity; and (6) stage of occlusion. The FP digital system appeared to be settled in a lower-dose rate for fluoroscopy (a factor of 6) and higher for dose per frame in cine (a factor of 3) in comparison with the image intensifier (II) system. There was a marked reduction of DAP when the FP technology was introduced.
Conclusion:
More extensive studies should be performed in the future so as to further investigate the influence of the FP detector in IC.
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Cardiac Catheterization IV: Nursing Management
Imaging Studies for Cardiovascular System V: CT
Biological Effects of Radiation