Effect of calcium tablets on interpretation of lumbar spine DXA scans
David L Kendler1, Gary M Kiebzak, Catherine G Ambrose
1Osteoporosis Research Centre, Vancouver, British Columbia, Canada.
Abstract:
Undissolved calcium (Ca) tablets in the gastrointestinal tract at the time of a dual-energy X-ray absorptiometry (DXA) scan could conceivably affect the accuracy of the lumbar spine bone mineral density (BMD) determination. We studied phantoms and volunteers to determine the effect of Ca tablets overlying bone, Ca tablets in the soft-tissue field, and Ca tablets overlapping both bone and soft tissue. For L1-4, a 500-mg Ca tablet taped to the phantom surface or the skin of volunteers resulted in