Hydrodynamic evaluation of kangaroo aortic valve matrices for tissue valve engineering
Related Experiment Video
Updated: Aug 8, 2026

An Ex Vivo Porcine Model for Hydrodynamic Testing of Experimental Aortic Valve Procedures and Novel Medical Devices
Published on: August 25, 2023
Kishan K Narine1, Katarzyna Kramm, Kris Dumont
1Department of Cardiac Surgery, University Hospital Ghent, Ghent, Belgium. kishan.narine@ugent.be
Abstract:
We evaluated the hydrodynamic performance of kangaroo aortic valve matrices (KMs) (19, 21, and 23 mm), as potential scaffolds in tissue valve engineering using a pulsatile left heart model at low and high cardiac outputs (COs) and heart rates (HRs) of 60 and 90 beats/min. Data were measured in two samples of each type, pooled in two CO levels (2.1 +/- 0.7 and 4.2 +/- 0.6 L/min; mean +/- standard errors on the mean), and analyzed using analysis of variance with CO level, HR, and valve type as fixed factors and compared to similar porcine matrices (PMs). Transvalvular pressure gradient (DeltaP) was a function of HR (P < 0.001) and CO (P < 0.001) but not of valve type (P = 0.39). DeltaP was consistently lower in KMs but not significantly different from PMs. The effective orifice area and performance index of kangaroo matrices was statistically larger for all sizes at both COs and HRs.