Related Experiment Video
Updated: Aug 8, 2026

The Flexible Rhino-Laryngoscope for Awake Nasotracheal Intubation
Published on: August 2, 2024
First clinical experience of tracheal intubation with the SensaScope, a novel steerable semirigid video stylet
P Biro1, U Bättig, J Henderson
1Department of Anaesthesiology, University Hospital Zurich, Zurich, Switzerland. peter.biro@usz.ch
Background:
Problems with tracheal intubation are a major cause of anaesthesia-related morbidity and mortality. Difficulty with tracheal intubation is primarily a consequence of failure to see the vocal cords with conventional direct laryngoscopy. We report our experience with use of the SensaScope for tracheal intubation in routine clinical practice.
Methods:
The SensaScope is a hybrid steerable semirigid S-shaped video stylet. Its handling and performance were assessed by anaesthetists with a minimum of 1 yr of experience. They performed four intubations each with the device in anaesthetized elective surgical patients. The view of the glottis with the Macintosh laryngoscope was compared with the view shown on the monitor by the SensaScope. The time taken to complete intubation, the final tracheal tube (TT) position and the degree of difficulty of the procedure were recorded.
Results:
Thirty-two patients were studied. All Macintosh Cormack and Lehane grade 3 patients were converted to grade 1 or 2 with the SensaScope. Mean intubation time was 25 (12) s and correct mid-tracheal TT cuff position was achieved in all cases. The degree of difficulty was 3.0 (1.8) on a numerical scale ranging from 0 to 10. All operators rapidly became familiar with the device and mastered its technique of use.
Conclusion:
The SensaScope is a reliable and effective device for tracheal intubation under vision of the normal airway. It has great potential to facilitate management of difficult airway situations in anaesthetized and paralysed patients.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy Suctioning I: Pre-Procedural Steps
Equipment Required
First, gather all necessary equipment: a sterile suction catheter, a sterile disposable container, sterile gloves, a towel or...
Tracheostomy Suctioning II: Procedure
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes: