Long-term tissue response to bioabsorbable poly-L-lactide and metallic screws: an experimental study
Harri Pihlajamäki1, Ole Böstman, Olli Tynninen
1Research Institute of Military Medicine, Central Military Hospital, Helsinki, Finland. harri.pihlajamaki@helsinki.fi <harri.pihlajamaki@helsinki.fi>
Bone
|June 6, 2006
Summary
Bioabsorbable poly-l-lactide (PLLA) and metallic screws showed similar long-term biocompatibility in rabbit bone. Both induced a bone walling-off response, with PLLA exhibiting slow degradation and fibrous tissue formation without adverse inflammatory reactions.
Area of Science:
- Biomaterials Science
- Orthopedic Surgery
- Tissue Engineering
Background:
- Late inflammatory reactions have been reported with bioabsorbable poly-l-lactide (PLLA) bone fixation devices.
- Understanding long-term tissue response to PLLA implants is crucial for clinical safety.
Purpose of the Study:
- To compare the long-term tissue response of bioabsorbable PLLA screws versus metallic screws in cancellous bone.
- To evaluate the biocompatibility and degradation of PLLA screws over extended follow-up periods.
Main Methods:
- Utilized rabbit femora models with PLLA and metallic screws, alongside intact bone controls.
- Employed radiography, microradiography, histology, histomorphometry, and oxytetracycline labeling.
- Follow-up assessments were conducted at 36 and 51 months.
Main Results:
- Both PLLA and metallic screws elicited a bone walling-off response.
- Connective tissue formed between bone and implant only in the PLLA group, with no significant difference in layer thickness over time.
- No significant differences in trabecular bone volume or osteoid formation were observed between screw types and controls.
- PLLA material diminished over time, indicating slow degradation accompanied by fibrous tissue formation and macrophage activity.
Conclusions:
- Bioabsorbable PLLA and metallic screws demonstrate comparable long-term biocompatibility in cancellous bone.
- PLLA screws undergo slow degradation with associated fibrous tissue formation, but without significant inflammatory foreign-body reactions.
- The observed tissue responses suggest inertness for both PLLA and metallic screws within the study's follow-up timeframe.