Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Cardiac output estimation with a new Doppler device after off-pump coronary artery bypass surgery
Rajesh Chand1, Yatin Mehta, Naresh Trehan
1Department of Anesthesiology and Critical Care, Escorts Heart Institute and Research Centre, New Delhi, India.
Objective:
To compare cardiac output (CO), stroke volume (SV), and cardiac index (CI) as estimated with a new, noninvasive Doppler device (Ultrasonic Cardiac Output Monitor [USCOM]; USCOM Ltd, Sydney, Australia) with those measured with the bolus thermodilution (TD) technique.
Design:
Prospective nonrandomized study.
Setting:
Postcardiac surgery recovery unit of a tertiary cardiac center.
Participants:
Fifty patients after off-pump coronary artery bypass (OPCAB) surgery.
Measurement And Main Results:
Both right-sided and left-sided CO were estimated with a USCOM continuous-wave (CW) Doppler device, and CO was determined with the bolus TD technique performed in triplicate. On comparing the right-sided CO, SV, and CI with those of TD, the mean bias was 0.03 L/min, 1.6 mL, and 0.02 L/min/m(2), respectively. The comparison of left-sided CO, SV, and CI with those of TD revealed a means bias of 0.14 L/min, 1.0 mL, and 0.08 L/min/m(2), respectively.
Conclusion:
This study showed excellent agreement between the values for CO, SV, and CI as determined with USCOM and TD. Since there was only 1 time period for CO estimation in each patient with both methods, the stability of this correlation needs to be further investigated over time.
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls between...