Related Experiment Video
Updated: Aug 8, 2026

Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Occupational radiation dose associated with Rb-82 myocardial perfusion positron emission tomography imaging
A Robert Schleipman1, Frank P Castronovo, Marcelo F Di Carli
1Division of Health Physics and Radiopharmacology, Department of Radiology, Brigham and Women's Hospital, Boston, MA 02115, USA. aschleipman@partners.org
Background:
We determined staff radiation dose during rest and stress rubidium 82 myocardial perfusion positron emission tomography (PET) imaging.
Methods And Results:
Patients received 1,587 +/- 163 MBq (42.9 +/- 4.4 mCi) Rb-82 during rest or pharmacologic stress. A pressurized ion chamber was used to monitor radiation exposure in 50 examinations. For comparison, staff exposure during pharmacologic stress in 20 other patients receiving 1,204 +/- 55.5 MBq (32.54 +/- 1.5 mCi) technetium 99m 2-methoxy isobutyl isonitrile (MIBI) was measured. For Rb-82 infusion and PET acquisition, the mean dose was 0.45 +/- 0.25 microSv (0.045 +/- 0.025 mrem). Exposure for routine stress testing at variable distances from the patient was equivalent to background. Similar exposure for pharmacologic stress testing through 7 minutes after injection of Tc-99m MIBI at variable distances was 1.075 +/- 0.32 microSv (0.108 +/- 0.03 mrem). However, exposure for stress tests starting 7 minutes after Rb-82 infusion at 0.5 m was estimated at 0.4 microSv (0.04 mrem). To determine the potential radiation dose for those responding to a medical emergency or otherwise in close proximity to a patient, we measured the mean cumulative dose at 0.5 m from 0 to 7 minutes of Rb-82 infusion, which resulted in 19.1 +/- 5.8 microSv (1.9 +/- 0.58 mrem).
Conclusions:
Radiation doses for all tasks during routine Rb-82 stress-rest PET are lower than measured Tc-99m MIBI values. However, the radiation dose in close proximity to the patient during or immediately after Rb-82 infusion can be considerably higher, underscoring the need for strict attention to source distance and contact times.
More Related Videos
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...