Effect of a posterior dynamic implant adjacent to a rigid spinal fixator
Thomas Zander1, Antonius Rohlmann, Nagananda K Burra
1Biomechanics Laboratory, Department of Orthopedics, Charité Campus Benjamin Franklin, Hindenburgdamm 30, 12203 Berlin, Germany.
Background:
A slightly degenerated disc adjacent to a segment that has to be fused is sometimes instrumented with a dynamic fixator. The dynamic implant is assumed to reduce disc loads at that level and to preserve disc function, thus inhibiting the progression of degeneration.
Methods:
A three-dimensional finite element model of the lumbar spine was used to study the effect of a dynamic implant on the mechanical behavior at the corresponding level. After studying a healthy lumbar spine for comparison, a rigid fixator and a bone graft were inserted at L2/L3. Healthy and degenerated discs were assumed at the adjacent level, i.e. L3/L4. An additional paired dynamic posterior fixator was then implemented at level L3/L4. Finally, the segment with the dynamic fixator was distracted to the height of a healthy disc. The loading cases of walking, extension, flexion and axial rotation were simulated.
Findings:
A dynamic implant reduces intersegmental rotation for walking, extension and flexion as well as facet joint forces for axial rotation at its insertion level. Intradiscal pressure is not markedly reduced by a dynamic implant. Moreover, there are no substantial differences between the mechanical behavior of rigid and dynamic fixators.
Interpretation:
Our model does not predict major differences in the mechanical effects between rigid and dynamic fixators despite the extreme assumption that a dynamic implant does not transfer moments. The results do not support the assumption that disc loads are significantly reduced by a dynamic implant. For axial rotation, however, dynamic fixation devices do reduce the force in the facet joint.