Direct Pressure Monitoring Accurately Predicts Pulmonary Vein Occlusion During Cryoballoon Ablation
Published on: February 26, 2013
Cryoablation of the pulmonary veins using a novel balloon catheter
Arthur Garan1, Amin Al-Ahmad, Teresa Mihalik
1New England Medical Center, Tufts University School of Medicine, Boston, MA, USA.
Introduction:
Pulmonary vein (PV) isolation has emerged as a promising technique for the treatment of patients with drug-refractory atrial fibrillation, however, the achievement of transmural lesions has remained a challenge. We evaluated the ability of a novel balloon-based cryogenic catheter system in achieving transmural lesions for PV isolation.
Methods:
Six pulmonary vein ostia from three excised ovine hearts and lungs were used in this study. The balloon catheter was deployed and positioned at the ostia of the PVs and a full 8-minute ablation was then performed, while the heart was bathed in a circulating bath of normal saline at 37 degrees. Thermocouples positioned on the endocardial (balloon surface-tissue interface) and epicardial surfaces of the ostia were used to determine whether transmural freezing was achieved.
Results:
The mean temperatures measured on the endocardial and epicardial tissue in six PV ablations were -38.8 +/- 6.9 degrees C and -10.0 +/- 7.5 degrees C, respectively. The average pulmonary vein thickness was 3.3 +/- 1.4 mm.
Conclusions:
A novel cryoablation balloon catheter is capable of achieving transmural freezing of the pulmonary vein. The catheter has promise for future clinical therapy of atrial fibrillation.