Related Experiment Video
Updated: Aug 7, 2026

Luminescence Lifetime Imaging of O2 with a Frequency-Domain-Based Camera System
Published on: December 16, 2019
Optical fibre dosemeter systems for clinical applications based on radioluminescence and optically stimulated
Carl Johan Marckmann1, Claus E Andersen, Marianne C Aznar
1Radiation Research Department, Risø National Laboratory, DK-4000 Roskilde, Denmark.
Abstract:
Optical fibre dosemeter systems based on radioluminescence and optically stimulated luminescence (OSL) from carbon-doped aluminium oxide (Al(2)O(3):C) crystals were developed for in vivo real-time dose rate and absorbed dose measurements in radiotherapy and mammography. A technique was also developed for making ultra-small dosemeter probes that can easily be placed inside patients in radiation treatment. These probes have shown excellent properties in both head and neck intensity-modulated radiation therapy treatment and in mammography. The dose-response of the OSL signal for the new optical fibre dosemeter system showed a repeatability within 0.15% at a dose level of 60 mGy when integrated over 100 s. The temperature dependence in the range 0-45 degrees C showed a reproducibility within 1.3% when the OSL signal was integrated over 100 s.
Related Concept Videos
Photoluminescence: Applications
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...