Transcatheter closure of perimembranous ventricular septal defects using amplatzer asymmetric ventricular septal

Robin J Pinto1, Bharat V Dalvi, Satyavan Sharma

  • 1Glenmark Cardiac Centre, Matunga (E), Mumbai, India.

Insights

The Amplatzer asymmetric ventricular septal defect occluder offers a promising nonsurgical option for perimembranous ventricular septal defects in children and adolescents. Short-term follow-up shows encouraging success rates and minimal complications.

Area of Science:

  • Cardiology
  • Pediatric Cardiology
  • Interventional Cardiology

Background:

  • Perimembranous ventricular septal defects (pmVSD) are common congenital heart defects.
  • Nonsurgical closure aims to correct these defects with less invasive methods.

Purpose of the Study:

  • To evaluate the efficacy and safety of the Amplatzer asymmetric ventricular septal defect occluder for transcatheter closure of pmVSD.
  • To assess the outcomes of this procedure in children and adolescents over an 18-month follow-up period.

Main Methods:

  • Twenty pediatric and adolescent patients with pmVSD underwent transcatheter closure using the Amplatzer device.
  • Patients received aspirin pre- and post-procedure.
  • Follow-up included clinical examination, ECG, and echocardiography at multiple time points up to 18 months.

Main Results:

  • The procedure was successful in 89.4% of attempted cases (17/19 patients).
  • A trivial residual shunt in 7 patients resolved by one month.
  • Complications included one anaphylactic reaction, three cases of right bundle branch block, one silent thromboembolism, and one case of hemolysis.

Conclusions:

  • The Amplatzer asymmetric ventricular septal defect occluder is a promising device for transcatheter closure of pmVSD in carefully selected pediatric and young adult patients.
  • Short-term follow-up demonstrates encouraging safety and efficacy, with no significant device-related complications observed during the study period.