Detection of V-type nerve agent degradation products at electrodes modified by PPy/PQQ using CaCl2 as supporting
Olga V Shulga1, Christopher Palmer
1Department of Chemistry, University of Montana, Missoula, MT 59812, USA. olga.shulga@mso.umt.edu
Abstract:
Electrochemical detection without derivatization was used to detect thiol-containing degradation products of V-type nerve agents. Electropolymerization of pyrrole was used for entrapment of the biocatalyst PQQ to produce a sensor. Various parameters which affect the detection processes such as the type of the supporting electrolyte used during electrodeposition and the thickness of the polypyrrole film were examined and optimized. Electocatalytic oxidation of thiols by the PPy/PQQ electrode was strongly affected by the presence of Ca2+ cations during electrodeposition of the PPy/PQQ. Cyclic voltammetry, linear sweep voltammetry and amperometry have been used for electrode characterization. Amperometric detection of the V-type nerve agent thiol degradation products 2-(dimethylamino)ethanethiol (DMAET) and 2-(diethylamino)ethanethiol (DEAET) was performed at 0.38 V. Linear calibration plots were observed for these compounds. The detection limits of 4.5 and 3 microM were obtained for DMAET and DEAET respectively, with sensitivities of 1.18 and 1.37 nA microM(-1) cm(-2).
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
Controlled-Potential Coulometry: Electrolytic Methods
The chosen potential ensures...