Related Experiment Video
Updated: Aug 7, 2026

High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Determination of gabapentin in human plasma using hydrophilic interaction liquid chromatography with tandem mass
Hye Young Ji1, Dong Won Jeong, Young Hoon Kim
1Drug Metabolism and Bioanalysis Laboratory, College of Pharmacy and Phytofermentation Research Center, Wonkwang University, Iksan 570-749, Korea.
Abstract:
A rapid, sensitive and selective method for the determination of gabapentin in human plasma was developed using hydrophilic interaction liquid chromatography/tandem mass spectrometry (HILIC/MS/MS). The devised method involved protein precipitation with acetonitrile followed by separation on an Atlantis HILIC silica column using an acetonitrile/ammonium formate mobile phase (100 mM, pH 3.0) (85:15, v/v). Analytes were detected using an electrospray ionization mass spectrometer in the multiple-reaction monitoring mode. The standard curve was linear (r = 1.000) over the concentration range of 50.0-10000 ng/mL. The lower limit of quantification for gabapentin was 50.0 ng/mL (ca. 20 pg gabapentin) using a 10-microL plasma sample. The coefficients of variation and relative errors for intra- and inter-assay at four QC levels (i.e., 50.0, 125, 750, and 7500 ng/mL) were 4.7 to 9.4% and -4.1 to 1.6%, respectively. Absolute and relative matrix effects for gabapentin and metformin were practically absent. Gabapentin and metformin recoveries were 98.5% and 99.0%, respectively. This method was successfully applied to a bioequivalence study of gabapentin in humans.