Sample to sample carryover: a source of analytical laboratory error and its relevance to integrated clinical
David A Armbruster1, Debra B Alexander
1Scientific Affairs, Abbott Laboratories, 100 Abbott Park Road, Abbott Park, IL 60046, USA. david.armbruster@abbott.com
Background:
Integrated systems that combine clinical chemistry and immunoassay analyzers are used routinely. Sample to sample carryover is an inherent risk and can cause erroneously high patient test results for immunoassays. IVD manufacturers and laboratories must be aware of this phenomenon and guard against it.
Methods:
We used a sample carryover protocol that directs the clinical chemistry module to process samples with very high immunoassay analyte concentrations followed by samples with very low concentrations for the same analyte. Low concentration samples were then tested by the immunoassay module to determine if the clinical chemistry module caused primary sample tube to primary sample tube carryover of the immunoassay analyte.
Results:
Sample carryover was assessed on the Abbott ci8200 for HBsAg, AFP, beta-hCG, and PSA. Observed HBsAg carryover met the design specification of <0.1 ppm. Carryover for the other analytes was <0.1 ppm or below the assay limit of detection.
Conclusions:
IVD manufacturers must design integrated systems to minimize primary specimen tube carryover and avoid analytical laboratory error that can impact patient safety. Carryover testing is difficult for clinical laboratories to perform in order to verify system performance. Laboratories must consider the potential for specimen carryover and its impact on results whether moving primary sample tubes between separate analyzers or using an integrated system.
More Related Videos
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Data Validation
Key parameters for method validation include:
Sample Handling
Samples should be transported carefully from collection points to the laboratory. They should be properly sealed and clearly labeled to prevent cross-contamination. To preserve the sample integrity, optimal temperature conditions during transport are essential. This could involve using...
Contaminants and Errors
Another key consideration is determining the appropriate number of samples required to...
Automated Microbial Diagnostics