A new calibration method that compensates for the effects of O2 and N2 on infrared CO2 analysers
Yongquan Tang1, Martin J Turner, A Barry Baker
1Department of Anaesthetics, University of Sydney, Royal Prince Alfred Hospital, Missenden Road, Camperdown, NSW 2050, Sydney, Australia.
Objective:
This study evaluates a method for calibrating mainstream CO(2) analysers in which CO(2) partial pressure (P (CO2)) is calculated as a function of the outputs of CO(2) and O(2) analysers.
Methods:
Three mass flow controllers were used to generate 25 different reference mixtures of O(2), N(2) and CO(2). Reference gas mixtures were combinations of P (CO2) = 2, 4, 6, 8, 10 kPa and O(2) partial pressure (P (O2)) = 10, 20, 40, 60, 80 kPa (balance N(2)). CO(2) and O(2) analyser data were fitted by a calibration equation which took into account the effects of oxygen partial pressure and nonlinearity of the CO(2) analyser. The calibration coefficients were tested in a separate validation data set with a variety of combinations of CO(2) and O(2).
Results:
Our new calibration method yields a standard deviation of CO(2) measurement error that is significantly lower than a CO(2)-only calibration method in the validation data set (0.54% versus 2.72%, P < 0.05). P (CO2) measurement errors produced by the single gas calibration equation are significantly correlated with P (O2) in both the calibration (R = -0.9906, P < 0.05) and validation data sets (R = -0.9642, P < 0.05), but the errors given by our new calibration equation are independent of P (O2) (R = -0.0364, NS, and R = -0.0305, NS, for calibration and validation data sets respectively). Calibration with only CO(2) cannot eliminate the error related to the collision broadening effect of O(2), which in our CO(2) analyser is approximately a 1% underestimation of P (CO2) for every 10 kPa (75 mmHg) increase in P (O2).
Conclusions:
This study shows that non-dispersive infrared CO(2) analyser readings can be substantially affected by background oxygen. This effect can be corrected for by calibrating the CO(2) analyser with gases containing known proportions of both CO(2) and O(2).
Related Concept Videos
IR Spectrometers
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Infrared (IR) Spectroscopy: Overview
Different compounds display unique properties due to their...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
IR Spectroscopy: Hooke's Law Approximation of Molecular Vibration
According to Hooke's law, the vibrational frequency is directly proportional to the...