Related Experiment Video
Updated: Aug 7, 2026

Gramicidin-based Fluorescence Assay; for Determining Small Molecules Potential for Modifying Lipid Bilayer Properties
Published on: October 13, 2010
Flow-injection solvent extraction without phase separation. Fluorimetric determination of thiamine by the thiochrome
A Alonso1, M J Almendral, M J Porras
1Departamento de Química Analítica, Nutrición y Bromatología, Faculty of Chemistry, University of Salamanca, Plaza de la Merced s/n, E-37008 Salamanca, Spain.
Abstract:
Two modes of liquid-liquid extraction in flow-injection systems were compared and applied to the fluorimetric determination of thiamine (Vitamin B(1)). The first included phase segmentation, but fluorescence was measured without phase separation. In this mode, thiamine was detected at concentrations higher than 8 microg/l with a linear application range of 30-2,000 microg/l, an R.S.D. of 1.9% (150 microg/l, n=10) and a sampling frequency of 60/h. In the second mode, a single segment of organic solution was injected into the aqueous stream and fluorescence was also measured without phase separation. Using this mode, concentrations of thiamine higher than 1 microg/l were detected, with a linear application range between 5 and 280 microg/l, an R.S.D. of 2.4% (150 microg/l, n=10) and a sampling frequency of 60/h. The two forms were applied to the analysis of thiamine in pharmaceuticals.