Three types of skin-surface thermometers: a comparison of reliability, validity, and responsiveness
Robert S Burnham1, Robert S McKinley, Daniel D Vincent
1Division of Physical Medicine and Rehabilitation, University of Alberta, Edmonton, Canada.
Objective:
To compare the reliability, validity, and responsiveness of a thermistor thermometer (thermistor) and two different infrared thermometers (one designed to measure tympanic temperature and one for skin temperature).
Design:
Reliability and validity were evaluated by making two separate measurements from the skin at identical spots of each hand, forearm, shoulder, thigh, shin, and foot in 17 healthy subjects. Intramuscular temperature was recorded at the hand and shin sites. Test-retest reliability was calculated using intraclass correlation for each instrument. Pearson correlation assessed the relationship between the skin and intramuscular temperatures at the hand and shin sites (validity). Each instrument's ability to measure temperature change (responsiveness) was assessed by measuring skin temperatures serially from 17 limbs of ten patients with complex regional pain syndrome undergoing intravenous regional sympathetic blockade. Responsiveness index values were calculated.
Results:
Reliability was strong and similar for each device (intraclass correlation: thermistor = 0.96, tympanic = 0.96, skin = 0.97), as was validity (r: thermistor = 0.90, tympanic = 0.92, skin = 0.92). Responsiveness was marginally better for the infrared skin device (responsiveness index: skin = 4.2, tympanic = 3.6, thermistor = 3.6).
Conclusions:
For the purposes of clinical electrodiagnostic laboratory and other physiatry applications, the performance of the infrared thermometers is equal to or superior to that of the traditionally used thermistor. All three devices are highly reliable and valid, whereas the infrared skin device is slightly more responsive. Infrared thermometers have the advantage of being quicker to operate and more portable.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Thermometers and Temperature Scales
As many physical properties depend on temperature, the variety of thermometers is...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...