[Precision and accuracy of "a pocket" pulse oximeter in Mexico City]
Luis Torre-Bouscoulet1, Elizabeth Chávez-Plascencia, Juan Carlos Vázquez-García
1Clínica de Suefio, Instituto Nacional de Enfermedades Respiratorias, Tlalpan, DF México.
Background:
Pulse oximeters are frequently used in the clinical practice and we must known their precision and accuracy. The objective was to evaluate the precision and accuracy of a "pocket" pulse oximeter at an altitude of 2,240 m above sea level.
Methods:
We tested miniature pulse oximeters (Onyx 9,500, Nonin Finger Pulse Oximeter) in 96 patients sent to the pulmonary laboratory for an arterial blood sample. Patients were tested with 5 pulse oximeters placed in each of the fingers of the hand oposite to that used for the arterial puncture. The gold standard was the oxygen saturation of the arterial blood sample.
Results:
Blood samples had SaO2 of 87.2 +/- 11.0 (between 42.2 and 97.9%). Pulse oximeters had a mean error of 0.28 +/- 3.1%. SaO2 = (1.204 x SpO2) - 17.45966 (r = 0.92, p < 0.0001). Intraclass correlation coefficient between each of five pulse oximeters against the arterial blood standard ranged between 0.87 and 0.99. HbCO (2.4 +/- 0.6) did not affect the accuracy.
Conclusions:
The miniature oximeter Nonin is precise and accurate at 2,240 m of altitude. The observed levels of HbCO did not affect the performance of the equipment. The oximeter good performance, small size and low cost enhances its clinical usefulness.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
07:51Evaluation of Capnography Sampling Line Compatibility and Accuracy when Used with a Portable Capnography Monitor
Published on: September 29, 2020
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring pulse
Pulse
The pulse serves as a clinical indicator...