Determination of cardiac volumes and mass with FLASH and SSFP cine sequences at 1.5 vs. 3 Tesla: a validation study
Lucy E Hudsmith1, Steffen E Petersen, Damian J Tyler
1University of Oxford Center for Clinical Magnetic Resonance Research, University of Oxford, Oxford, UK.
Purpose:
To compare cardiac cine MR imaging using steady state free precession (SSFP) and fast low angle shot (FLASH) techniques at 1.5 and 3 T, and to establish their variabilities and reproducibilities for cardiac volume and mass determination in volunteers. To assess the feasibility of SSFP imaging in patients at 3 T and to determine comparability to volume data acquired at 1.5 T.
Materials And Methods:
Ten healthy volunteers underwent cardiac magnetic resonance imaging using SSFP and segmented gradient-echo FLASH, using both a 1.5 and a 3 T MR system on the same day. Ten patients with impaired left ventricular (LV) function were also studied at both field strengths with SSFP.
Results:
For both SSFP and FLASH, field strength had no effect on the quantification of LV and right ventricular (RV) volumes, mass, or function (P > or = 0.05 for field strength for all parameters). At both 1.5 and 3 T, SSFP yielded smaller LV mass (e.g., at 3 T 109 +/- 30 g vs. 142 +/- 37 g; P = 0.011) and larger LV volume (e.g., at 3 T end-diastolic volume 149 +/- 37 mL vs. 133 +/- 31 mL at 5 T; P = 0.041) measurements than FLASH. In patients with reduced LV function, all volume and mass measurements were again similar for SSFP sequences at 1.5 vs. 3 T. In volunteers and patients, measurement variabilities for LV parameters were small for both field strength and sequences, ranging between 3.7% and 10.7% for mass.
Conclusion:
Compared to 1.5 T, cardiac cine MR imaging at 3 T, using either FLASH or SSFP sequences, is feasible and highly reproducible. Field strength does not have an influence on quantification of cardiac volume or mass, but the systematic overestimation of LV mass and underestimation of LV volume by FLASH compared to SSFP is present at both 1.5 and 3 T. Normal values for cardiac volumes and mass established at 1.5 T can be applied to scans obtained at 3 T.
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...