Dynamic cine-CT angiography for the evaluation of the thoracic aorta; insight in dynamic changes with implications
B E Muhs1, K L Vincken, J van Prehn
1Department of Vascular Surgery, University Medical Center Utrecht, The Netherlands.
Objective:
Thoracic aneurysm preoperative imaging is performed using static techniques without consideration of normal aortic dynamics. Improved understanding of the native aortic environment into which thoracic endografts are placed may aid in device selection. It is unclear what comprises normal thoracic aortic pulsatility. We studied these phenomena dynamically using ECG-gated 64-slice CTA.
Methods:
Maximum diameter and area change per cardiac cycle was measured at surgically relevant anatomic thoracic landmarks in ten patients; 1.0 cm proximal and distal to the subclavian artery, 3.0 cm distal to the subclavian artery, and 3.0 cm proximal to the celiac trunk. Data was acquired using a novel ECG-gated dynamic 64-slice CT scanner during a single breath hold with a standard radiation dose and contrast load. Eight gated data sets, covering the cardiac cycle were reconstructed, perpendicular to the central lumen.
Results:
There is impressive change in both maximum diameter and area in the thoracic aorta during the cardiac cycle. Mean maximum diameter changes of greater than 10% are observed in the typical sealing zones of commercially available endografts corresponding to diameter increases of up to 5mm. Aortic area increases by over 5% per cardiac cycle.
Conclusions:
ECG-gated dynamic CTA with standard radiation dose is feasible on a 64-slice scanner and provides insight into (patho) physiology of thoracic aortic conformational changes. Clinicians typically oversize thoracic endografts by 10%. With aortic pulsatility resulting in diameter changes of up to 17.8%, the potential exists for endograft undersizing, graft migration, intermittent type I endoleak, and poor patient outcome. Furthermore, aortic pulsatility is not evenly distributed, and non-circular stentgraft designs should be considered in the future since aortic distension in the aneurysm neck is not evenly distributed.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Aneurysm II: Clinical Manifestations and Diagnostic Studies
Imaging Studies for Cardiovascular System V: CT
Aortic Regurgitation II: Clinical Features and Diagnostic Tests
Acute Coronary Syndrome III: Diagnostic Studies
Aneurysm III: Interprofessional Care